简介概要

管式炉加热红外吸收法测定锰矿中低含量硫

来源期刊:冶金分析2011年第3期

论文作者:武素茹 王虹 姚传刚 马德起

文章页码:68 - 71

关键词:锰矿; 碳硫分析仪; 低硫; 末线时间

Key words:manganese ore; carbon/sulfur analyzer; low content sulfur; endline time

摘    要:采用高温管式炉加热红外吸收法,建立了锰矿中低硫含量检测的方法,重点探讨了仪器参数中的末线时间、比较器水平及分析稳定比较器等参数的选取对结果的影响,得到了测硫的最佳试验条件。结果表明,该方法用于锰矿石中硫的检测,线性范围为0.002%~0.032%,方法检出限为0.001%,相对标准偏差小于5%。

Abstract: The determination method of low content sulfur in manganese ore was established by high-temperature tube furnace heating coupled with infrared absorption method.The effect of instrument parameters on the determination results were discussed,including endline time,comparator level and analytical stability comparator.The optimal experimental conditions for determination of sulfur were obtained.This method has been applied to the detection of sulfur in manganese ore.The results showed that the linear range was 0.002%-0.032%,the detection limit was 0.001%,and the relative standard deviation(RSD) was smaller than 5%.

详情信息展示

管式炉加热红外吸收法测定锰矿中低含量硫

武素茹1,王虹2,姚传刚2,马德起2

(1.天津市天津出入境检验检疫局
2.天津出入境检验检疫局,天津 300456)

摘 要:采用高温管式炉加热红外吸收法,建立了锰矿中低硫含量检测的方法,重点探讨了仪器参数中的末线时间、比较器水平及分析稳定比较器等参数的选取对结果的影响,得到了测硫的最佳试验条件。结果表明,该方法用于锰矿石中硫的检测,线性范围为0.002%~0.032%,方法检出限为0.001%,相对标准偏差小于5%。

关键词:锰矿; 碳硫分析仪; 低硫; 末线时间

<上一页 1 下一页 >

相关论文

  • 暂无!

相关知识点

  • 暂无!

有色金属在线官网  |   会议  |   在线投稿  |   购买纸书  |   科技图书馆

中南大学出版社 技术支持 版权声明   电话:0731-88830515 88830516   传真:0731-88710482   Email:administrator@cnnmol.com